LS Software User Guide
116
Changing Molecular Weight Settings
Add Molecular Weight Standard to a Library
Edit Molecular Weight Standard to a Library
Remove Molecular Weight Standard to a Library
Copy Molecular Weight Standard to a Library
Calibration of molecular weight (MW) involves associating a known standard with one or more lanes in
the image. This allows Rf values to be calibrated to molecular weight values. The software allows the
user to employ several different standards per gel. To help with analysis, a library of molecular weight
standards is provided in the software. Standards can be added, edited, and deleted from the library using
the following instructions.
Add Molecular Weight Standard to the Library
First identify all lanes and bands in the image.
Select the lane to calibrate by clicking on it. [ Edit Lanes and Bands ]
Select 1D Analysis Action Tab > Lane Profile menu and click Calibrate lane.
In the Step 1: Select MW standard window, click Add. A new MW standard window opens. In
this window, provide the new Name of the standard, Group, and Unit Type.
Click Add. Now the numerical values of the standard can be entered. After entering the numerical
value, click on Add again for as many values as desired.
Click OK on the right side. The first window appears again with the new standard entered.